B-1000 Infant Incubator

  • Model: B-1000

Price: Price Unavailable

STANDARD COMPOSITION:


●Servo Control Mode: air temp;
● Instant Parameter Displaying: set temp, air temp and heat power rate;
● Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air temp override,fan failure,system failure;
● >37℃ Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function;
● Hood with dual slope: 6 hand-ports; left & right cabinets, top-shelf;
● RS-232 Output Connector (Use for network system ).

 

 

 

TECHNICAL SPECIFICATION
 

Specifications of hood
B-3000
B-2000
B-1000
B-800
Mattress to Hood
450 mm
450 mm
450 mm
410 mm
Observation side of hood
Two incline
Two incline
Two incline
Two incline
Operate port
6 (2 Iris ports)
6 (2 Iris ports)
6 (2 Iris ports)
4 (2 Iris ports)
Tubing grommets
4
4
4
2
Front access panel
826×310 mm
826×310 mm
826×310 mm
826×280 mm
Mattress Tray Size
630×360 mm
630×360 mm
630×360 mm
630×360 mm
Technical Specifications:
 
Power requirement
≤400VA
≤400VA
≤400VA
≤400VA
Temperature control mode
Air and skin servo control
Air and skin servo control
Air servo control
Air servo control
Air Temp Control Range
20℃37℃
20℃37℃
20℃37℃
20℃37℃
Temperature override mode to
20℃39℃
20℃39℃
20℃39℃
----
Skin Temp Control Range
30℃37℃
30℃37℃
----
----
Temperature override mode to
30℃38℃
30℃38℃
----
----
Skin temp sensor accuracy
≤0.3℃
≤0.3℃
----
----
Humidity display range
5%99%
----
----
----
Temperature fluctuation
≤0.5℃
≤0.5℃
≤0.5℃
≤0.5℃
Mattress temperature uniformity
≤0.8℃
≤0.8℃
≤0.8℃
≤0.8℃
Temperature rise time
≤30 min
≤30 min
≤30 min
≤30 min
Noise level within hood environment
≤55dB
≤55dB
≤55dB
≤55dB
Alarms
 
Power failure alarm
Yes
Yes
Yes
Yes
Fan failure alarm
Yes
Yes
Yes
Yes
Sensor failure alarm
Yes
Yes
Yes
Yes
High air temperature alarm
+1.5℃
+1.5℃
+1.5℃
+1.5℃
Low air temperature alarm
-2.5℃
-2.5℃
-2.5℃
-2.5℃
High skin temperature alarm
+1.0℃
+1.0℃
----
----
Low skin temperature alarm
-1.0℃
-1.0℃
----
----
Over air temperature alarm
≤38.0℃; ≤40.0℃
≤38.0℃; ≤40.0℃
≤38.0℃; ≤40.0℃
≤38.0℃
Over skin temperature alarm
≤38.0℃; ≤39.0℃
≤38.0℃; ≤39.0℃
≤38.0℃; ≤39.0℃
----
System failure alarm
Yes
Yes
Yes
----
Alarm silence function
Yes
Yes
Yes
----
Operational condition
 
Environment temperature
+20℃+30℃
+20℃+30℃
+20℃+30℃
+20℃+30℃
Environment relative humidity
5%99%
5%99%
5%99%
5%99%
Environment air velocity of flow
<0.3m/s
<0.3m/s
<0.3m/s
<0.3m/s
Others
 
RS-232 Connector
Yes
Yes
Yes
-----
Rail
Yes
Yes
Yes
-----
Infant bed tilt adjusting
±10° step-less
±10° step-less
±10° step-less
-----
Mattress tray Max. load
10kg
10kg
10kg
10kg
Transfusion pole Max. load
2kg
2kg
2kg
2kg
Cabinet Max. load
6kg
6kg
6kg
----
Packing size (2 cartons)
100×58×80 cm
100×58×80 cm
100×58×80 cm
100×58×98 cm
Gross Weight
120kg
120kg
120kg
83 kg

 

 

 

 

OPTIONAL PART

               

YC      Infant Bed adjustment,  adjust tile ±10°with step-less.

SC Double wall canopy.

 

           

GL    Blue light Phototherapy Unit,  LED digital timer,   display therapy time and total time.

 GB Top shelf for put other device.

              

HL  Protect machine, to prevent hit in moving. 

YEC  Infant scale, to weighing infant weight.

 

 

 

 

SGL

Blue light Phototherapy Unit with up and below,

LED digital timer, display therapy time and total time.

Up phototherapy unit used blue tube light source,

below phototherapy unit used blue LED light source.

 

 

SPECIFICATION

POWER
Model B-1000-1E...............................................................................................................................110 / 120 Vac, 50Hz / 60Hz
Model B-1000-2E...............................................................................................................................220 / 240 Vac, 50Hz / 60Hz
Power Input .................................................................................................................................................................... ≤400VA

Refer to IEC60601-1, this equipment belongs to ClassⅠ, Type BF.  Its operation is continuous.

ALARM
Power Alarm
Power Failure.................................................Activates if primary power to the Incubator 
fails or 
the power wire is accidentally disconnected from the wall receptacle.
Fan Alarm
Fan Failure.........................................................................................................................Activates if Fan motor malfunctions.
Sensor Alarm
Air Sensor Failure...............................................................................................Air Sensor malfunction, refer unit to service.

Deviation Alarm
High Air Temperature.......................................................Activated if displayed Air Temperature is 1.5℃above set point.
Low Air Temperature........................................................Activated if displayed Air Temperature is 2.5℃below set point.
Over Temperature Alarm
Over Air Temperature......................................................Activated when displayed Air Temperature was approach 38℃ for set temperature ≤37℃, or approach 40℃ for set temperature >37℃ in Air Mode.
System Alarm
System Failure ..................................................................................Activated if Temperature Control System malfunctions.

 


TEMPERATURE CONTROL RANGE AND CORRELATIVE SPECIFICATIONS
Air Temperature........................................20℃~37℃,for set temperature ≤37℃, 37℃~39℃,for set temperature >37℃
Temperature Rise Time ..................................................................................................................................................... ≤30min
Temperature Variability....................................................................................................................................................... ±0.5℃
Temperature Uniformity......................................................................................................................................................... ≤1℃

ENVIRONMENT REQUIREMENT
Environment Temperature
Operation Temperature Range.............................................................................................................................+20℃ ~ +30℃
Store Temperature Range......................................................................................................................................-40℃ ~ +55℃
Environment Humidity
Operation Humidity Range............................................................................................................................................30%~75%
Store Humidity Range...................................................................................................................................................10%~100%
Atmospheric Pressure Range........................................................................................................................... 700hPa~1060hPa
OTHER SPECIFICATION
Noise Level Within Hood Environment.................................................................... ≤55dB(Environment noise below 45dB)
Carbon Dioxide (CO2) Level within the Hood.......................................................Less than 0.5% when a mixture of CO2 in air is delivered at 750ml/min at a point 10cm above the center of the mattress.
Air Velocity Over Mattress..................................................................................................................................less than 0.35m/s
Infant Bed Max Load...................................................................................................................................................................10kg
Transfusion Shelf Max Load........................................................................................................................................................2kg
Cabinet Max Load ....................................................................................................................................................................... 6kg

FUNCTION OF “”SILENCE KEY
Make the alarm to be silence for 10 minutes by press this key, following:
Deviation Alarm
Sensor Alarm
Fan Alarm

FACTORY DEFAULT SETTINGS
Temperature Control Mode.......................................................................................................................Air Temperature Mode
Air Set Temperature.....................................................................................................................................................................32℃
Temperature Units............................................................................................................................................................................℃

NOTE: Open Access Panel or Access Door or the use of supplies or other equipment within the Incubator, which can alter the air flow pattern, may effect temperature uniformity, temperature variability, the correlation of the Incubator temperature reading to center mattress temperature and skin temperature.